Straight Handpiece

PM 1:1 Irrig

This last-generation handpiece is the result of our ongoing research and development work. PM 1:1 is designed to withstand the stresses resulting from repeated lateral pressure, while maintaining exceptional precision. You will appreciate the absence of vibration, quiet operation and tried and tested bur locking system.

Reference:

PM 1:1 external irrigation 1600052-001

PM 1:1 Irrig
 
PM 1:1 external irrigation
TypeStraight handpiece
Ratio1:1 (direct)
CouplingE-type as per ISO 3964
Locking systemLocking ring
Bur dimensionBur shank Ø 2.35 mm
Cleaning / sterilisationThermal washer / steam autoclave up to 135°C (275°F)
IrrigationInterchangeable sleeves with or without irrigation tube
Warranty2 to 3  years 1
Reference PM 1:1 external irrigation 1600052-001
PM 1:1 external irrigation
Type Straight handpiece
Ratio 1:1 (direct)
Coupling E-type as per ISO 3964
Locking system Locking ring
Bur dimension Bur shank Ø 2.35 mm
Cleaning / sterilisation Thermal washer / steam autoclave up to 135°C (275°F)
Irrigation Interchangeable sleeves with or without irrigation tube
Warranty 2 to 3  years 1
Reference PM 1:1 external irrigation 1600052-001

1 2-year standard warranty and 1-year optional warranty available through Bien-Air’s PlanCare extended warranty program.

Maintenance of contra-angles

PREPARATION
A.

REMOVING THE INSTRUMENT

  • Remove the rotary instrument from the chuck mechanism
  • Disconnect the contra-angle from the motor
B.

PRE-DISINFECTION

  • External disinfection

Do not immerse the instrument in a disinfectant bath.

CLEANING
A.

CLEAN UNDER RUNNING WATER AT <38°C

B.

REMOVING DIRT/DEPOSITS

  • Carefully remove all traces of dirt or deposits from the nozzles using the Bien-Air cleaning wire
  • Dry the nozzles with compressed air then dry the rest of the instrument with a cloth

Do not immerse in an ultrasonic cleaner

C.

INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET

  • Shake the Spraynet before use
  • Use the Spraynet attachment 1
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second
  • Clean the exterior of the contra-angle with a cloth soaked in Spraynet
DISINFECTION
A.

DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer's instructions for use)

Do not immerse the instrument in a disinfectant bath

OR
B.

THERMAL DISINFECTION (follow the manufacturer's instructions for use)

LUBRICATION
A.

LUBRICATION WITH LUBRIFLUID

  • Shake the Lubrifluid before usen
  • Use the Lubrifluid attachment 1
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second
  • Then, run an operating test for 30 seconds at low speed to remove any surplus lubricant
  • Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out
STERILISATION
A.

PLACE THE CONTRA-ANGLE IN A STERILISATION BAG

B.

STERILISE IN ACCORDANCE WITH THE MANUFACTURER'S INSTRUCTIONS

  • Instruments bearing the "135°C" can be sterilised in an autoclave

Only sterilise instruments which have been cleaned and lubricated